SıCAKLıK SENSöRü AOIP CALYS 50 MULTIFONKSIYONEL PROSES KALIBRATöRü-SıCAKLıK KALIBRATöRü

Multifonksiyonel Proses Kalibratörü-Sıcaklık Kalibratörü



Marka : AOIP
Model : CALYS 50
Ürün Kodu : CALYS50
SıCAKLıK SENSöRü AOIP CALYS 50 MULTIFONKSIYONEL PROSES KALIBRATöRü-SıCAKLıK KALIBRATöRü
Özellikler

Sensor Range (Input and Output) Resolution Accuracy / 1 year (Measurement) Accuracy / 1 year (Simulation)
Pt50 (α = 3851) -220°C to +1200°C 0.01°C 0.012% RDG + 0.06°C 0.014% RDG + 0.18°C
Pt100 (α = 3851) -220°C to +1200°C 0.01°C 0.012% RDG + 0.05°C 0.014% RDG + 0.12°C
Pt100 (α = 3916) -200°C to +510°C 0.01°C 0.012% RDG + 0.05°C 0.014% RDG + 0.12°C
Pt100 (α = 3926) -210°C to +850°C 0.01°C 0.012% RDG + 0.05°C 0.014% RDG + 0.12°C
Pt200 (α = 3851) -220°C to +600°C 0.01°C 0.012% RDG + 0.12°C 0.014% RDG + 0.33°C
Pt500 (α = 3851) -220°C to +1200°C 0.01°C 0.012% RDG + 0.07°C 0.014% RDG + 0.18°C
Pt1000 (α = 3851) -220°C to +1200°C 0.01°C 0.012% RDG + 0.05°C 0.014% RDG + 0.08°C
Ni100 (α = 618) -60°C to 180°C 0.01°C 0.012% RDG + 0.03°C 0.014% RDG + 0.08°C
Ni120 (α = 672) -40°C to +205°C 0.01°C 0.012% RDG + 0.03°C 0.014% RDG + 0.08°C
Ni1000 (α = 618) -60°C to +180°C 0.01°C 0.012% RDG + 0.03°C 0.014% RDG + 0.08°C
Cu10 (α = 427) -50°C à to 150°C 0.01°C 0.012% RDG + 0.18°C 0.014% RDG + 0.1°C
Cu50 (α = 428) -50°C to +150°C 0.01°C 0.012% RDG + 0.06°C 0.014% RDG + 0.15°C

Resistive probes measurements in 2, 3 or 4 wires: automatic recognition of number of connected wires, with indication on screen
Accuracies are given for 4-wire mounted probes
Take into account particular error of temperature sensor used and implementation conditions
Measuring current: 0.25 mA (Measurement) or from 0.1 to 1 mA (Emission)
Temperature coefficient: < 10% of accuracy /°C

Thermocouples: Measurement and simulation

Type Input range Resolution Accuracy / 1 year (Measurement) Output range Resolution Accuracy / 1 year (Simulation)
K -250 to -200°C
-200 to -120°C
-120 to +0°C
0°C to +1372°C
0.2°C
0.1
0.05°C
0.05°C
0.8°C
0.25°C
0.1°C
0.013% RDG + 0.08°C
-240 to -50°C
-50 to -0°C
+0°C to +1372°C
0.2°C
0.1
0.05°C
0.6°C
0.1°C
0.013% RDG + 0.08°C
T -250 to -200°C
-200 to -120°C
-120 to -50°C
-50 to +400°C
0.2°C
0.05°C
0.05°C
0.05°C
0.7°C
0.25°C
0.1°C
0.013% RDG + 0.08°C
-240 to -100°C
-100 to -0°C
+0 to +400°C
0.2°C
0.05°C
0.05°C
0.4°C
0.1°C
0.013% RDG + 0.08°C
J -210 to -120°C
-120 to +0°C
+0 to +1200°C
0.05°C
0.05°C
0.05°C
0.25°C
0.09°C
0.013% RDG + 0.07°C
-210 to -0°C
+0 to +1200°C
0.05°C
0.05°C
0.2°C
0.013% RDG + 0.07°C
E -250 to -200°C
-200 to -100°C
-100 to -0°C
+0 to +1000°C
0.1°C
0.05°C
0.05°C
0.05°C
0.45°C
0.15°C
0.07°C
0.013% RDG + 0.05°C
-240 to -100°C
-100 to +40°C
+40 to 1000°C
0.1°C
0.1°C
0.05°C
0.25°C
0.1°C
0.013% RDG + 0.05°C
R -50 to +150°C
+150 to +550°C
+550 to 1768°C
0.5°C
0.2°C
0.1°C
0.8°C
0.013% RDG + 0.35°C
0.013% RDG + 0.2°C
-50 to +350°C
+350 to +900°C
+900 to 1768°C
0.5°C
0.2°C
0.1°C
0.5°C
0.013% RDG + 0.35°C
0.013% RDG + 0.2°C
S -50 to +150°C
+150 to +550°C
+550 to +1768°C
0.5°C
0.2°C
0.1°C
0.80°C
0.013% RDG + 0.35°C
0.013% RDG + 0.25°C
-50 to +120°C
+120 to +450°C
+450 to +1768°C
0.5°C
0.2°C
0.1°C
0.8°C
0.013% RDG + 0.35°C
0.013% RDG + 0.25°C
B +400 to +900°C
+900 to +1820°C
0.2°C
0.1°C
0.013% RDG + 0.4°C
0.013% RDG + 0.2°C
+400 to +850°C
+850 to +1820°C
0.2°C
0.1°C
0.013% RDG + 0.4°C
0.013% RDG + 0.2°C
U -200 to +660°C 0.05°C 0.15°C -200 to +600°C 0.05°C 0.15°C
L -200 to +900°C 0.05°C 0.2°C -200 to +900°C 0.05°C 0.2°C
C -20 to +900°C
+900 to 2310°C
0.1°C
0.1°C
0.25°C
0.013% RDG + 0.15°C
-20 to +900°C
+900 to 2310°C
0.1°C
0.1°C
0.25°C
0.013% RDG + 0.15°C
N -240 to -190°C
-190 to -110°C
-110 to +0°C
+0 to +1300°C
0.2°C
0.1°C
0.05°C
0.05°C
0.5°C
0.15°C
0.08°C
0.013% RDG + 0.06°C
-240 to -190°C
-190 to -110°C
-110 to +0°C
+0 to +1300°C
0.2°C
0.1°C
0.05°C
0.05°C
0.3°C
0.15°C
0.08°C
0.015% RDG + 0.06°C
Platinum -100 to +1400°C 0.05°C 0.3°C -100 to +1400°C 0.05°C 0.3°C
Mo +0 to +1375°C 0.05°C 0.013% RDG + 0.06°C +0 to +1375°C 0.05°C 0.013% RDG + 0.06°C
NiMo/NiCo -50 to +1410°C 0.05°C 0.013% RDG + 0.30°C -50 to +1410°C 0.05°C 0.013% RDG + 0.3°C

Thermocouples: D and G, for specifications, refer to the instruction manual (Available on request)
Accuracy is given for reference @ 0°C.
When using the internal reference junction (except couple B) add an additional uncertainty of 0.2 °C at 0 °C.
It is possible (thermocouple B excepted) to choose by programming the cold junction localization: External at 0°C, internal (temperature compensation of instrument’s terminals) or manually entered.
Temperature coefficient: <10% of accuracy /°C
Display unit: °C and F

Specifications and performances in process @23°C ±5°C

DC current: Measurement

With or without loop supply

Range Res. Accuracy / 1an Rin
0-20 mA 1 µA 0.0175% RDG + 2 µA < 25 Ω
4-20 mA 1 µA 0.0175% RDG + 2 µA < 25 Ω
±50 mA 1 µA 0.0175% RDG + 2 µA < 25 Ω

Temperature coefficient: < 10 ppm/°C beyond reference domain
Loop supply: 24 V ±10%
HART® compatibility: Input impedance Rin = 280 Ω
Display with linear or quadratic scaling

DC voltage: Measurement

Range Res. Accuracy / 1an Rin
100 mV 1 µV 0.013% RDG + 3 µV > 10 MΩ
1 V (1) 10 µV 0.013% RDG + 20 µV > 10 MΩ
10 V 100 µV 0.015% RDG + 200 µV > 1 MΩ
50 V 1 mV 0.015% RDG + 2 mV > 1 MΩ

Temperature coefficient < 7 ppm/°C beyond reference domain
(1) Specification domain: -0.8 V to +1 V 

Frequency, counting: Measurement

Range Resolution Accuracy / 1an
20 kHz 0.01 Hz 0.005% RDG

Temperature coefficient < 5 ppm/°C beyond reference domain
Scale unit: Pulse / min and Hz
Trigger level: 1 V
Measurement on frequency signals or dry contacts
Counting will be performed on defined time or infinite time.

Resistance: Measurement

Range Resolution Accuracy / 1an
400 Ω 1 mΩ 0.012% RDG + 10 mΩ
400 Ω 10 mΩ 0.012% RDG + 100 mΩ

Resistance measurement in 2, 3 or 4 wires: automatic recognition of number of connected wires, with indication on screen
Accuracies are given for 4-wire mounted probes
Temperature coefficient: < 7 ppm/°C beyond reference domain
Open circuit terminal voltage: < 10 V
Continuity test: Open circuit for R > 1000 Ω and closed circuit for R < 1000 Ω
Measurement current: 0.25 mA

DC current: Emission

With or without loop supply

Range Resolution Accuracy / 1an
24 mA 1 µA 0.0175% RDG + 2 µA
4-20 mA 1 µA 0.0175% RDG + 2 µA
0-20 mA 1 µA 0.0175% RDG + 2 µA

Temperature Coefficient < 10 ppm/°C beyond reference domain
Settling time: < 5 ms
Specifications given for CALYS configurations in:
- Active mode (+24V ON) 1  Meter in passive mode (+24 V OFF)
- Passive mode (+24 V OFF) 1 Meter in active mode (+24 V ON)
Pre-programmed steps
                               0%       25%      50%      75%     100%
4-20 mA linear            4          8         12         16        20
0-20 mA linear            0          5         10         15        20
4-20 mA quad             4          5          8         13        20
0-20 mA quad             0        1.25        5       11,25      20
4-20 mA valves    3.8-4—4.2               12               19, 20, 21

DC voltage: Emission

Range Res. Accuracy / 1an Min load
100 mV 1 µV 0.013% RDG + 3 µV 1 kΩ
2 V 10 µV 0.013% RDG + 20 µV 2 kΩ
20 V 100 µV 0.015% RDG + 200 µV 4 kΩ

Noise: 3 ppm (for 0.1 Hz to 10 Hz) and 5 ppm (for 10 Hz to 100 Hz)
Temperature coefficient: < 7 ppm/°C beyond reference domain
Settling time: < 5 ms

Frequency, pulse: Emission

Range Resolution Accuracy / 1an
1000 Hz 0.01 Hz 0.005% RDG
10 kHz 0.1 Hz 0.005% RDG

Temperature coefficient < 5 ppm/°C beyond reference domain 

Resistance: Emission

Range Res. Accuracy / 1an Nota
Iext
40 Ω 1 mΩ 0.014% RDG + 3 mΩ
0.014% RDG + 10 mΩ
10 mA
1 mA
400 Ω 10 mΩ 0.014% RDG + 20 mΩ
0.014% RDG + 30 mΩ
10 mA
1 mA
4000 Ω 100 mΩ 0.014% RDG + 300 mΩ 0.1 / 1 mA

Temperature coefficient: < 5 ppm/°C beyond reference domain
Current establishing time: < 1 ms
Iext: Current received by the calibrator

Further features

Scaling in measurement and simulation modes Scaling allows process signals to be displayed in % of FS or in all other units. This function also allows sensors to be corrected after a calibration.
Relative measurement The features allows the following :
  • Programming a reference value different from the one of the instrument (NUL function).
  • Subtracting of constant value by measuring or programming it from a measured value (TARE function).
Simulation menu Simulation value is set by entering value on keypad or by changing the specific digit with the cursor.
Square root In current measurement and simulation, this function allows taking into account a quadratic signal coming from transmitter of type ΔP.
Statistical functions Continuous display of average, minimum and maximum value of the signal under monitoring, as well as number of measurements.
Ramps generation Starting, ending and length time values of simple or cyclic ramps can be set to do simulation.
Number of ramps can also be adjusted in case of cyclic ramps for any signals.
Steps simulation 2 modes are available:
  • Program mode: Starting value, number of steps and the length time have to be set
  • Manual mode: User has about a hundred preset values

In current simulation, user will have some additional preset values in function of range and according to 0%, 25%, 50%, 75% and 100% from selected gauge. Choice is done between gauges:
0-20 mA: linear or quadratic
4-20 mA: linear or quadratic

Synthesizer With 100 values manually set, CALYS 50 enables users to draw a generation curve.
Transmitter function CALYS 50 is able to be used as a transmitter. Measurement input is copied on the output with scaling.

 

General specifications

Size 210 x 110 x 50 mm
Weight 900 g
Display 240 x 320 pixel liquid crystal graphical display with backlite and contrast control
Display of result as table of values or trend curve
Power supply 230 V ± 10 %, 50/60 Hz
Battery Type: Lithium-ion
Charging Time: 3 hours
Lifetime: 8 hours
Communication ports USB

Environmental specifications

Reference range 23°C ± 5°C (RH: 45 to 75 % w/o condensing)
Operating reference range -10 to 50°C (RH: 20 to 80 % w/o condensing)
Limit operating range -15°C to +55°C (RH: 10 to 80 % w/o condensing) (70% at 55°C)
Storage temperature limits -30°C to +60°C
Maximum height 0 to 2000 m
IP protection IP54 according to EN60529

 Safety specifications

Protections
  • Electronic protection up to 250 V for ‘voltage’ wires
  • Fuse protection for ‘current’ wires
  • Protection against ‘current’ circuit breaking during inductive resistance measurements
Class In accordance with EN 61010-1
Category II, pollution 2
Rated voltage 60 V
Chocks and vibrations EN 61010-1
EMC conformity Immunity:
  • EN 61000-4-2
  • EN 61000-4-3
  • EN 61000-4-5
  • EN 61000-4-6
  • EN 61000-4-11

EN 61000-4-4
Conducted and radiated emissions:

  • EN 55022, class B
  • EN 61000-3-2
  • EN 61000-3-3